Chapter 13 301
SCPI Command Reference
E5061A/E5062A commands
13. SCPI Command
Reference
:CONT:HAND:F
Syntax :CONTrol:HANDler:F[:DATA] <numeric>
Description Outputs data to output port F (port A + port B) of the handler I/O. Data is outputted as
16-bit binary using A0 as LSB and B7 as MSB. (No query)
For details about the handler I/O, see Chapter 10.
Parameters
If the specified parameter is out of the allowable setup range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
Example of use
10 OUTPUT 717;":CONT:HAND:F 511"
Related commands :CONT:HAND:A on page 295
:CONT:HAND:B on page 295
Equivalent key No equivalent key is available on the front panel.
<numeric>
Description Output data
Range 0 to 65535
Resolution 1