Agilent Technologies E5071C Betriebsanweisung Seite 363

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 551
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 362
Chapter 13 363
SCPI Command Reference
E5061A/E5062A commands
13. SCPI Command
Reference
:MMEM:STOR:STYP
Syntax :MMEMory:STORe:STYPe {STATe|CSTate|DSTate|CDSTate}
:MMEMory:STORe:STYPe?
Description Selects the contents saved when saving the instrument state into a file with the
:MMEM:STOR command.
Parameters
Query response {STAT|CST|DST|CDST}<newline><^END>
Example of use
10 OUTPUT 717;":MMEM:STOR:STYP CDST"
20 OUTPUT 717;":MMEM:STOR:STYP?"
30 ENTER 717;A$
Related commands :MMEM:STOR on page 353
Equivalent key
[Save/Recall] - Save Type - State Only|State & Cal|State & Trace|All
Description
STATe
Specifies the save of the measurement conditions
*1
only.
CSTate
(preset value)
Specifies the save of the measurement conditions
*1
and the calibration
state.
DSTate
Specifies the save of the measurement conditions
*1
and the formatted
data array.
CDSTate
Specifies the save of the measurement conditions
*1
, the calibration
state, and the formatted data array.
*1.For details about the measurement conditions that can be saved, refer to the Users
Guide.
Seitenansicht 362
1 2 ... 358 359 360 361 362 363 364 365 366 367 368 ... 550 551

Kommentare zu diesen Handbüchern

Vikiehs 04 Jan 2024 | 13:20:06

urenrjrjkvnm