Agilent Technologies E5071C Betriebsanweisung Seite 61

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 551
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 60
Chapter 4 61
Performing a Calibration
Calibration
4. Performing a Calibration
preceding command completes.
As shown in Table 4-1, the data required to calculate the calibration coefficients differs
depending on the selected calibration type.
In the data section in the table, the letter m (for example, 1, a) represents the measurement
data at port m; m-n (for example, 1-2, a-b) represents the measurement data between
response port m and stimulus port n. You can omit data enclosed in brackets.
Table 4-1 Calibration Types and Required Data
Calibration type
Selected ports are
enclosed in parentheses.
Data
OPEN SHORT LOAD THRU Isolation
Response
OPEN (a) a Not required [a] Not required Not required
SHORT (a) Not required a [a] Not required Not required
THRU (a-b) Not required Not required Not required a-b [a-b]
Enhanced Response (a-b) b b b a-b [a-b]
Full 1-Port (a) a a a Not required Not required
Full 2-Port (a-b) a, b a, b a, b a-b, b-a [a-b], [b-a]
Seitenansicht 60
1 2 ... 56 57 58 59 60 61 62 63 64 65 66 ... 550 551

Kommentare zu diesen Handbüchern

Vikiehs 04 Jan 2024 | 13:20:06

urenrjrjkvnm