Chapter 13 213
SCPI Command Reference
E5061A/E5062A commands
13. SCPI Command
Reference
:CALC{1-4}:CORR:EDEL:TIME
Syntax :CALCulate{[1]|2|3|4}[:SELected]:CORRection:EDELay:TIME <numeric>
:CALCulate{[1]|2|3|4}[:SELected]:CORRection:EDELay:TIME?
Description Sets the electrical delay time of the active trace of channel 1 (:CALC1) to channel 4
(:CALC4) (specified with the :CALC{1-4}:PAR{1-4}:SEL command).
Parameters
If the specified parameter is out of the allowable setup range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
Query response {numeric}<newline><^END>
Example of use
10 OUTPUT 717;":CALC1:CORR:EDEL:TIME 0.2"
20 OUTPUT 717;":CALC1:CORR:EDEL:TIME?"
30 ENTER 717;A
Related commands :CALC{1-4}:PAR{1-4}:SEL on page 286
Equivalent key
[Scale] - Electrical Delay
<numeric>
Description Electrical delay time
Range -10 to 10
Preset value 0
Unit s (second)