Agilent Technologies E5071C Betriebsanweisung Seite 378

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 551
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 377
378 Chapter 13
SCPI Command Reference
E5061A/E5062A commands
:SENS{1-4}:CORR:COLL:CKIT:ORD:OPEN
Syntax :SENSe{[1]|2|3|4}:CORRection:COLLect:CKIT:ORDer:OPEN <numeric 1>,<numeric 2>
:SENSe{[1]|2|3|4}:CORRection:COLLect:CKIT:ORDer:OPEN? <numeric 1>
Description For the calibration kit selected for channel 1 (:SENS1) to channel 4 (:SENS4), selects the
standard used for the open measurement of the specified port.
Parameters
If the specified parameter is out of the allowable setup range, an error occurs and the
command is ignored.
Query response {numeric 2}<newline><^END>
Example of use
10 OUTPUT 717;":SENS1:CORR:COLL:CKIT:ORD:OPEN 1,2"
20 OUTPUT 717;":SENS1:CORR:COLL:CKIT:ORD:OPEN? 1"
30 ENTER 717;A
Related commands :SENS{1-4}:CORR:COLL:CKIT on page 375
Equivalent key
[Cal] - Modify Cal Kit - Specify CLSs - Open - Port 1|Port 2
<numeric 1> <numeric 2>
Description Port number Standard number
Range 1 to 2 1 to 21
Resolution 1 1
Seitenansicht 377
1 2 ... 373 374 375 376 377 378 379 380 381 382 383 ... 550 551

Kommentare zu diesen Handbüchern

Vikiehs 04 Jan 2024 | 13:20:06

urenrjrjkvnm