Chapter 13 471
SCPI Command Reference
E5061A/E5062A commands
13. SCPI Command
Reference
:STAT:QUES:RLIM:PTR
Syntax :STATus:QUEStionable:RLIMit:PTRansition <numeric>
:STATus:QUEStionable:RLIMit:PTRansition?
Description Sets the value of the positive transition filter of the Questionable Ripple Limit Status
Register.
Parameters
Note that bit 5 to 15 cannot be set to 1.
Query response {numeric}<newline><^END>
Example of use
10 OUTPUT 717;":STAT:QUES:RLIM:PTR 16"
20 OUTPUT 717;":STAT:QUES:RLIM:PTR?"
30 ENTER 717;A
Related commands :STAT:QUES:RLIM? on page 465
:STAT:QUES:RLIM:NTR on page 470
Equivalent key No equivalent key is available on the front panel.
<numeric>
Description Value of the positive transition filter
Range 0 to 65535
Preset value Varies depending on the upper limit setting for channel/trace
number.
Resolution 1