Agilent Technologies E5071C Betriebsanweisung Seite 395

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 551
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 394
Chapter 13 395
SCPI Command Reference
E5061A/E5062A commands
13. SCPI Command
Reference
:SENS{1-4}:CORR:COLL:ECAL:ERES
Syntax :SENSe{[1]|2|3|4}:CORRection:COLLect:ECAL:ERES <numeric 1>,<numeric 2>
Description Executes enhanced response calibration of channel 1 (:SENS1) to channel 4 (:SENS4)
using the ECal (Electrical Calibration) module. If you execute this command when the
ECal module is not connected, an error occurs and the command is ignored. (No query)
Parameters
For each parameter, you must specify a different port number. If you specify the same port
number for 2 or more parameters, an error occurs and the command is ignored.
Example of use
10 OUTPUT 717;":SENS1:CORR:COLL:ECAL:ERES 1,2"
20 OUTPUT 717;"*OPC?"
30 ENTER 717;A
Equivalent key [Cal] - ECal - Enhanced Response - 2-1(S21)|1-2(S12)
:SENS{1-4}:CORR:COLL:ECAL:ISOL
Syntax :SENSe{[1]|2|3|4}:CORRection:COLLect:ECAL:ISOLation[:STATe] {ON|OFF|1|0}
:SENSe{[1]|2|3|4}:CORRection:COLLect:ECAL:ISOLation[:STATe]?
Description For channel 1 (:SENS1) to channel 4 (:SENS4), turns ON/OFF the isolation measurement
when executing Ecal (Electrical Calibration).
Parameters
Query response {1|0}<newline><^END>
Example of use
10 OUTPUT 717;":SENS1:CORR:COLL:ECAL:ISOL ON"
20 OUTPUT 717;":SENS1:CORR:COLL:ECAL:ISOL?"
30 ENTER 717;A
Related commands :SENS{1-4}:CORR:COLL:ECAL:SOLT1 on page 397
:SENS{1-4}:CORR:COLL:ECAL:SOLT2 on page 397
<numeric 1> <numeric 2>
Description Response port number Stimulus port number
Range 1 to 2 1 to 2
Resolution 1 1
Description
ON or 1 Turns ON the isolation measurement.
OFF or 0 (preset value) Turns OFF the isolation measurement.
Seitenansicht 394
1 2 ... 390 391 392 393 394 395 396 397 398 399 400 ... 550 551

Kommentare zu diesen Handbüchern

Vikiehs 04 Jan 2024 | 13:20:06

urenrjrjkvnm