Agilent Technologies ESG Spezifikationen Seite 70

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 170
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 69
Performance Tests ESG Family Signal Generators
3. Phase Modulation Accuracy and Distortion
4-12 Calibration Guide
3. Phase Modulation Accuracy and Distortion
These automated tests verify the phase modulation accuracy and distortion specifications.
Accuracy is verified directly with a measuring receiver, and distortion is verified by
measuring the demodulated output from the measuring receiver, using the audio analyzer.
The audio analyzer provides a more accurate verification with better resolution than using
the measuring receiver alone.
Recommended Equipment
HP/Agilent 8902A Measuring Receiver Option 010
HP/Agilent 8903B Audio Analyzer
Equipment Setup
Figure 4-10 Accuracy & Distortion for Phase Modulation Tests Setup
Seitenansicht 69
1 2 ... 65 66 67 68 69 70 71 72 73 74 75 ... 169 170

Kommentare zu diesen Handbüchern

Keine Kommentare