Seitenansicht 133
Adjustments ESG Family Signal Generators
Burst Audio Path Gain Calibration
5-28 Calibration Guide
Burst Audio Path Gain Calibration
Description
This test adjusts the scaling of the linear burst audio path until a -0.99 Vdc input to the
EXT1 connector results in a 40 dB drop in power relative to 0.00 Vdc input when linear
burst is activated. The AM 1 DAC value is changed until this result is achieved and its
value is stored as a calibration array.
Required Test Equipment
HP/Agilent 3458A Digital Multimeter
HP/Agilent 33120A 15 MHz Function/Arbitrary Waveform Generator
HP/Agilent 8563E Spectrum Analyzer
Procedure
Figure 5-24 Burst Audio Path Gain Calibration Setup
1. Connect the equipment as shown above.
2. Preset all of the equipment.
3. Follow the instructions as they appear on the controller’s display.