Agilent Technologies ESG Spezifikationen Seite 131

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 170
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 130
ESG Family Signal Generators Adjustments
Low Frequency (LF) Output Calibration
Calibration Guide 5-25
Low Frequency (LF) Output Calibration
Description
The LF output provides a calibrated audio frequency signal. This test is used to set the
full-scale amplitude of the LF-OUT DAC on the reference board. When the calibration is
complete, the voltage on the LF output should be equal to the input voltage, plus or minus
1 mV. The resulting DAC values are then stored in the signal generators firmware.
Required Test Equipment
HP/Agilent 3458A Digital Multimeter
Procedure
Figure 5-21 Low Frequency (LF) Output Calibration Setup
1. Connect the equipment as shown above.
2. Preset all of the equipment.
3. Follow the instructions as they appear on the controller’s display.
Seitenansicht 130
1 2 ... 126 127 128 129 130 131 132 133 134 135 136 ... 169 170

Kommentare zu diesen Handbüchern

Keine Kommentare