Operation Verification ESG Family Signal Generators
Verification Procedures
2-8 Calibration Guide
Configure the Measuring Receiver
1. Reset the measuring receiver.
2. Set the measuring receiver to FM mode.
3. Turn on Peak+ detector.
4. Turn on the 300 Hz high-pass filter.
5. Turn on the 3 kHz low-pass filter.
Measure the Deviations
1. Set the signal generator to the frequencies shown in Table 2-5, “FM Accuracy,” on
page 2-19.
2. Record the deviations measured and compare them to the limits shown in Table 2-5.
6. AM Accuracy Check
Connect the Test Equipment
Figure 2-5 AM Accuracy Equipment Setup
Configure the Signal Generator
1. Preset the signal generator to normal (factory-defined) preset conditions.
2. Press
AM > AM Off On. The AM annunciator is displayed.
3. Press
AM Rate. Enter 1 using the numeric keypad and press the kHz terminator softkey.
4. Press
Amplitude. Enter 7 using the numeric keypad and press the dBm terminator
softkey.
5. Check that the MOD ON annunciator is displayed. If it is not, press Mod On/Off.
6. Press
RF On/Off. The RF ON annunciator is displayed.