74 1100 Series DAD and MWD Reference Manual
3 Troubleshooting and Test Functions
UV Lamp Voltage
The UV lamp anode voltage is missing.
The processor continually monitors the anode voltage across the lamp during
operation. If the anode voltage falls below the lower limit, the error message is
generated.
Probable Causes
• Defective UV lamp or non-Agilent lamp.
• Defective detector main board.
• Defective power supply.
Suggested Actions
✔ Exchange the UV lamp.
✔ Exchange the detector main board.
✔ Exchange the power supply.