1100 Series DAD and MWD Reference Manual 145
Repairing the Detector 4
7 Remove the detector main board. Place the board on the ESD kit.
8 In most cases the RFI spring plate remains on the interface connectors of
the board. Carefully remove the spring plate and place it back into its
position in the instrument before installing a new board.
9 On the new board check the switch setting of address switch S1, see
“Setting the 8-bit Configuration Switch” on page 261.
10 Install the new detector main board and reconnect the connectors. Assure
that the board is fitted correctly into the board recess in the rear panel.
11 Refit the screws at the REMOTE and GPIB connectors and the nuts of the
ANALOG connectors.
12 Reinstall the top foam section, top cover and front cover, see “Replacing the
Foam and the Top Cover” on page 175.
13 Replace detector into the stack and reconnect the cables.
14 Check the firmware revision of the module. If the firmware revision is older
than the current firmware revision of the module, update the firmware
using the standard firmware update procedure, see “Replacing the
Detector’s Firmware” on page 180.
NOTE
An incorrect switch setting (for example, TEST/BOOT) may cause the module to turn in a
basic mode (yellow or red flashing status light). In such a case turn off the module, reset
the address switches, and turn on the module again.
NOTE
If a new detector main board is installed, update the serial number information of the
detector in the user interface, see procedure below.
NOTE
The G1315B DAD and the G1365B MWD require firmware revision 4.x or above (main and
resident).