162 1100 Series DAD and MWD Reference Manual
4 Repairing the Detector
Replacing the Holmium Oxide Filter Motor
When required If defective
Tools required Screwdriver POZI 1 PT3
hexagonal key 3 mm
Parts required Filter motor assembly G1315-68700 (includes filter lever G1315-45001 and
spring 1460-1510)
For others see “Holmium Oxide Filter” on page 197
NOTE
If the filter motor was removed, the filter lever should not be reused. Always use a new
filter lever to assure correct fit on the filter motor shaft.
Preparations for this procedure:
• Turn off the lamp.
• Switch off the detector, and disconnect the power cable.
• Remove detector from stack and place it on working
bench.
• Remove front cover, top cover and top foam section, see
“Removing the Top Cover and Foam” on page 140.
• Remove optical unit, see “Removing the Optical Unit
(G1315A/G1365A)” on page 154 or “Removing the
Optical Unit (G1315B/G1365B)” on page 156.
1 Place the optical unit on your bench.
Kommentare zu diesen Handbüchern