Agilent Technologies 1100 Series Wartungshandbuch Seite 120

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 318
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 119
120 1100 Series DAD and MWD Reference Manual
4 Repairing the Detector
Notes
The fittings at the flow cell body are special types for
low dead volumes and not compatible with other
fittings.
When retightening the fittings, make sure that they
are carefully tightened (handtight plus 1/4 turn with
a wrench). Otherwise damage of the flow cell body
or blockage may result.
2 After replacing the outlet capillary, fix it handtight first.
Then do a 1/4 turn with a 4-mm wrench.
3 To replace the inlet capillary, use a 4-mm wrench for the
fitting.
4 Unscrew the cell body from the heat exchanger and the
heat exchanger from the clamp unit.
Outlet capillary
Inlet capillary
Seitenansicht 119
1 2 ... 115 116 117 118 119 120 121 122 123 124 125 ... 317 318

Kommentare zu diesen Handbüchern

Keine Kommentare