122 1100 Series DAD and MWD Reference Manual
4 Repairing the Detector
Nano Flow Cell - Replacing or Cleaning
When required If parts are contaminated or leaky.
Tools required Screwdriver POZI 1 PT3
Two 1/4 inch wrenches for capillary connections
Parts required For parts identification refer to “Nano Flow Cells” on page 192 (80 nl and
500 nl).
NOTE
For details refer to the technical that comes with the nano-flow cell kit.
NOTE
The quartz block can be cleaned with alcohol. DO NOT touch the inlet and outlet windows
at the quartz block.
1 Disconnect the capillaries from the capillary holder and
remove the flow cell.
2 Unscrew the cell body from the holder.