156 1100 Series DAD and MWD Reference Manual
4 Repairing the Detector
Removing the Optical Unit (G1315B/G1365B)
When required For all repairs inside the optical unit
Tools required Screwdriver POZI 1 PT3
Parts required Optical unit G1315-69002 (exchange assembly) or
Individual parts depending on the following procedures
NOTE
With the introduction of the G1315B/G1365B DAD/MWD the appearance of the z-plane
(front metal plane) has been changed (no holes any more) and an isolation seal between
the optical unit and the z-plane. Therefore, some of the figures may still show the old
appearance of the G1315A/G1365A. For G1315A/G1365A detectors refer to “Removing the
Optical Unit (G1315A/G1365A)” on page 154.
Preparations for this procedure:
Turn off the lamp.
Switch off detector, and disconnect power cable.
Remove detector from stack and place it on working
bench.
Remove front cover, top cover and top foam section, see
“Removing the Top Cover and Foam” on page 140.
1 .Loosen the screw on the left side of the z-plane and the
two screes near the hydraulic connection.