124 1100 Series DAD and MWD Reference Manual
4 Repairing the Detector
7 Replace the cell seal assembly onto the quartz body.
Always use a new seal assembly to exclude damage
during disassembling.
8 Slide the quartz body completely into the cell body to the
front stop (use for example a toothpick).
9 Insert the flow cell capillaries and tighten them
fingertight. Use the wrench and torque adapter as
described on page 126 and tighten the fittings alternately.
10 Reassemble the flow cell body to the holder.