1100 LC Modules Control Manual - 78 - 03-914947-30:Rev.6
Define the center wavelength in the Wl field and the bandwidth in
the Bw field. The signal values contained within the bandwidth
are bunched and give one value for the center wavelength.
NOTE: For more details about the reference wavelength settings, refer to your Agilent
Technologies 1100 Diode Array Detector technical documentation.
Signal Program Parameter Group
Program changes the wavelength and/or the bandwidth for each
signal and/or its reference during runs.
Use the following buttons to edit the table:
to add a line to the table.
to delete a line in the table.
to clear the entire table.
Others Parameter Group
Check Automatic Lamp Stop to stop automatically the detector
lamp at the end of the run.
Select the Peakwidth value in the scrolling list. This parameter
influences the peak detection. A response time and a data rate
are associated with each peakwidth value. The available values
are:
<0.01 min (Response Time <0.1 sec) (DataRate=20 Hz)
>0.01 min (Response Time =0.2 sec) (DataRate=20 Hz)
>0.02 min (Response Time =0.5 sec) (DataRate=10 Hz)
>0.05 min (Response Time =1 sec) (DataRate=5 Hz)
>0.1 min (Response Time =2 sec) (DataRate=2.5 Hz)
>0.2 min (Response Time =4 sec) (DataRate=1.25 Hz)
>0.4 min (Response Time =8 sec) (DataRate=0.62 Hz)