1100 LC Modules Control Manual - 6 - 03-914947-30:Rev.6
NOTE: For more details about the configuration of a GPIB_Interface or an NI488 bus,
refer to the Galaxie Chromatography Data System Installation Guide.
Connecting the Agilent LC 1100 modules to Galaxie
The Galaxie Chromatography Data System uses GPIB to
communicate with the Agilent 1100 modules.
For each Agilent 1100 module controlled by Galaxie, a switch
located on the rear panel of the instrument must be configured
appropriately. For all the modules, the first three switches from
left must be turned off (Down Position) for control with Galaxie as
for control with HP-CHEMSTATION. Changing the position of
these switches must be done with module switched-off.
1 2 3 4 5 6 7 8
The module where the GPIB cable is connected should be the
detector of the HPLC chain (as recommended by Agilent). This
module is called the Master module. The rest of the modules of
the HPLC chain communicate through the Master module via
CAN cables. The GPIB address entered in each module in the
Galaxie Configuration Manager must be the GPIB address of the
master module.
Up (on)
Down (off)