Agilent Technologies 1100 Series Installationsanleitung Seite 168

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 184
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 167
1100 LC Modules Control Manual - 168 - 03-914947-30:Rev.6
Self-Test Test
The Self-Test test consists in running the following individual
tests consecutively: Filter Test, Slit Test, Dark Current Test,
Wavelength Calibration test, Intensity test, Holmium Test,
spectral flatness test and ASTM noise test if checked before
starting the Self-test.
Select Single Self-test or Repetitive Self-Test to run the test once
or repetitively. Note that if one of the tests fails for the first cycle,
the second cycle will not be started.
The ASTM noise test determines the detector baseline noise (at
254 nm). It must be performed while pumping water at 1 mL/min.
Results are displayed at screen and also in the maintenance
audit trail:
Seitenansicht 167
1 2 ... 163 164 165 166 167 168 169 170 171 172 173 ... 183 184

Kommentare zu diesen Handbüchern

Keine Kommentare