DisplayPort Test Implementation Group 36 Agilent J-BERT N4903B & 90000A/X/Q Series Sink Draft2
new waveform Press Clear Screen to clear the previous acquisition date, then the Single
button to acquire a new waveform.
Note: If a N4916B is De-emphasis Signal Converter is part of the setup ISI can be
reduced by applying a post cursor, see E appendix.
Eye Diagram Display
1. To create Eye diagram, select Analysis > Serial Data
a. Select Serial Data Wizard.
b. Use same clock recovery set in Jitter Wizard.
Figure 33: Serial Data Wizard Settings…
c. Click Next.
d. Click through Thresholds and Time Interval Measurements screens with the default settings.
e. In Real Time Eye, select Turn on real time eye display and Use a color graded display.