DisplayPort Test Implementation Group 15 Agilent J-BERT N4903B & 90000A/X/Q Series Sink Draft2
9.2 RJ Calibration
1. Calibrate Rj(rms) using D24.3 pattern with JMD using constant clock at TP3_EQ. For RBR, measure
at Rj at TP3.
a. In Pattern > Select Pattern Menu:
• Select User Pattern from File.
• Select Browse
• Select Pattern:
C:\N4903B\Pattern\Demo\DisplayPort\D24.3.ptrn
Figure 7: J-BERT N4903B D24-3_Clock Pattern
b. In the J-BERT N4903B Jitter > Jitter Setup Menu, disable all jitter terms accept RJ
Figure 8: J-BERT N4903B HBR2 RJ Adjustment