Agilent Technologies DisplayPort Bedienungsanleitung Seite 16

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 56
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 15
Agilent Technologies, Inc.
DisplayPort Test Implementation Group 16 Agilent J-BERT N4903B & 90000A/X/Q Series Sink Draft2
c. In the oscilloscope, using the Analysis > Serial Data > Clock Recovery menu, change the
clock recovery from 2nd Order PLL to Constant Frequency.
Figure 9: Constant Frequency CDR for RJ and SJ Calibration
d. Alternate between adjusting RJ on the J-BERT N4903B user interface and press the Single
Acquisition Button on the Oscilloscope until the RJ is calibrated to the following values.
For RBR
For HBR
For HBR2
Adjust RJ to 8.1mUI RMS
Repeat measurement at least 3
times and average the results.
Adjust RJ to 13.5mUI RMS
Repeat measurement at least 3
times and average the results.
Adjust RJ to 16.7mUI RMS,
200mUI Pk-Pk (16.7 RMS x
12.0)
Repeat measurement at least 3
times and average the results.
Seitenansicht 15
1 2 ... 11 12 13 14 15 16 17 18 19 20 21 ... 55 56

Kommentare zu diesen Handbüchern

Keine Kommentare