Agilent Technologies DisplayPort Bedienungsanleitung Seite 17

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 56
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 16
Agilent Technologies, Inc.
DisplayPort Test Implementation Group 17 Agilent J-BERT N4903B & 90000A/X/Q Series Sink Draft2
Figure 10: HBR2 RJ Measurement
Note: In this example, the measured RJ value is 16.4mUI RMS. The measured RJ is
multiplied by 12.0 to reach the 10
-9
BER value in the specification to use for the
correction in 9.3.1. In this case, this is 16.4 x 12.0 = 196.8mUI.
9.3 SJ Calibration
9.3.1 SJ
FIXED
Calibration
1. For HBR2 only, Calibrate SJ
FIXED
using D24.3 pattern with JMD using constant clock at TP3_EQ.
a. In the J-BERT N4903B Jitter > Jitter Setup Menu, disable all Jitter terms accept PJ 1
b. Set the frequency in PJ 1 to 200MHz.
Figure 11: J-BERT N4903B HBR2 SJ
FIXED
Adjustment
Seitenansicht 16
1 2 ... 12 13 14 15 16 17 18 19 20 21 22 ... 55 56

Kommentare zu diesen Handbüchern

Keine Kommentare