Programming Guide 1-69
ESG Family Signal Generators Preparing for Use
Programming the Status Register System
The Data Questionable Calibration Event Register latches transition events from the
condition register as specified by the transition filters. Event registers are destructive
read-only. Reading data from an event register will clear the content of that register.
To query the event register, send the command
STATus:QUEStionable:CALibration[:EVENt]?
Figure 1-40.
The Data Questionable Calibration Status Group also contains a Data Questionable
Calibration Event Enable Register. This register lets you choose which bits in the Data
Questionable Calibration Event Register will set the summary bit (bit 8 of the Data
Questionable Condition register) to 1. Send the
STATus:QUEStionable:CALibration:ENABle <num> command where <num> is the sum of
the decimal values of the bits you want to enable. For example, to enable bit 9 and bit 3 so
that whenever either of those bits is set to 1, the Data Questionable Calibration summary
bit of the Data Questionable Condition Register will be set to 1, send the command
STAT:QUES:CAL:ENAB 520 (512 + 8). The command
STATus:QUEStionable:CALIbration:ENABle? returns the decimal value of the sum of
the bits previously enabled with the STATus:QUEStionable:CALibration:ENABle <num>
command.