Programming Guide 2-55
ESG Family Signal Generators Programming Commands and Examples
:POWer Subsystem
RF Output Reference Power State
:POWer:REFerence:STATe ON|OFF|1|0
:POWer:REFerence:STATe?
This command sets the operating state of the signal generator’s RF output reference. The
choices are On (1) or Off (0).
*RST Value: Off
RF Output Start Power
:POWer:STARt <val><unit>
:POWer:STARt?
This command sets the first operating power level in swept power applications. The choices
for the variables <val> and <unit> are
−135.00 dBm to +20 dBm.
*RST Value:
−135.00 dBm
RF Output Stop Power
:POWer:STOP <val><unit>
:POWer:STOP?
This command sets the last operating power level in swept power applications. The choices
for the variables <val> and <unit> are
−135.00 dBm to +20 dBm.
*RST Value:
−135.00 dBm