1-8 Programming Guide
Preparing for Use ESG Family Signal Generators
Setting up the Equipment for Remote Operation
Connecting the Interface
1. Attach the male end of the RS-232 cable to the signal generator’s rear-panel
AUXILIARY INTERFACE connector.
2. Attach the female end of the RS-232 cable to the null modem.
3. Using a 5-mm nut driver, remove both standoffs from the female-to-female adapter.
4. Connect one end of the modified adapter to the null modem and the other end to the
selected port on the computer.
For personal computers, make the connection to the COM2 RS-232 port. COM1 is
acceptable if available.
For UNIX workstations, connect to the /dev/tty00 port. Alternatively, /dev/tty01 may
be used if it is available.
Configuring the Interface
1. Set the signal generator’s baud rate.
Press
Utility > GPIB/RS-232 > RS-232 Baud Rate. The default RS-232 baud rate is 19200.
Use baud rates 19200 or lower only. Press the appropriate softkey to adjust the signal
generator’s baud rate to the baud rate of your personal computer or UNIX workstation.
2. Set the signal generator’s handshake.
Press Utility > GPIB/RS-232 > RS-232 Pace. To determine the hardware operating
parameters, you need to know whether DSR (data set ready) and CTS (clear to send) are
active during communication with the controller, and the baud rate expected by the
controller. Set the signal generator’s handshake accordingly.
3. Set the signal generator’s RS-232 echo.
Press
Utility > GPIB/RS-232 > RS-232 Echo Off On to the desired operating state for your
configuration.
4. Set the signal generator’s remote language.
Press
Utility > GPIB/RS-232 > Remote Language. The only remote language supported by
RS-232 is SCPI (standard commands for programmable instruments).
If SCPI is not highlighted, press SCPI to select it.
Verifying the RS-232 System Interface
The following command verifies that the serial interface functions. With the equipment set
up as described in the previous sections, clear and reset the controller. Execute the
following command: OUTPUT "*IDN?"
The signal generator should return a string similar to the following, depending on model:
<instrument model name and number>, US37040098 B.03.00
If not, verify that the RS-232 parameters are set correctly (see “Setting Up the Interface
for GPIB Operation” on page 1-5) and that the interface cable is properly connected.