Command Reference
Pulse Modulation Subsystem ([:SOURce])
Chapter 4290
:PULM:INTernal[1]:PERiod
Supported PSG-A Series
[:SOURce]:PULM:INTernal[1]:PERiod <val><unit>|UP|DOWN
[:SOURce]:PULM:INTernal[1]:PERiod?
This command sets the period for the internally generated pulse modulation source.
*RST
+
2.00000000E
−
006
Range 70nS–42S
Key Entry
Pulse Period
Remarks If the entered value for the pulse period is equal to or less than the
value for the pulse width, the pulse width changes to a value that is
less than the pulse period.
Refer to “:PULM:INTernal[1]:PERiod:STEP[:INCRement]” for setting
the value associated with the UP and DOWN choices.
:PULM:INTernal[1]:PERiod:STEP[:INCRement]
Supported PSG-A Series
[:SOURce]:PULM:INTernal[1]:PERiod:STEP[:INCRement] <val><unit>
[:SOURce]:PULM:INTernal[1]:PERiod:STEP[:INCRement]?
This command sets the step increment for the internal pulse period.
*RST +1.00000000E-006
Range 10nS–42S
Key Entry N/A
Remarks The value set by this command is used with the UP and DOWN choices
for the pulse period command. Refer to “:PULM:INTernal[1]:PERiod”
for more information.
Choices <val><unit> UP DOWN