Command Reference
Status Subsystem (:STATus)
Chapter 4 193
:QUEStionable:CALibration:CONDition
Supported PSG-A Series
:STATus:QUEStionable:CALibration:CONDition?
This command returns the decimal sum of the bits in the Data Questionable Calibration
Condition Register. For example, if the DCFM or DC
Φ
M zero calibration fails (bit 0), a
value of 1 is returned.
*RST N/A
Range 0–32767
Key Entry N/A
Remarks Refer to Data Questionable Calibration Status Group” on page 133
and “Data Questionable Calibration Condition Register” on page 134
for more information.
The data in this register is continuously updated and reflects the
current conditions.
:QUEStionable:CALibration:ENABle
Supported PSG-A Series
:STATus:QUEStionable:CALibration:ENABle <value>
:STATus:QUEStionable:CALibration:ENABle?
This command determines what bits in the Data Questionable Calibration Event
Register will set the calibration summary bit (bit 8) in the Data Questionable Condition
Register.
The variable <value> is the sum of the decimal values of the bits that you want to
enable.
*RST N/A
Range 0–32767
Key Entry N/A
Remarks Refer to Data Questionable Calibration Status Group” on page 133
and “Data Questionable Calibration Event Enable Register” on
page 135 for more information.