Command Reference
Power Subsystem ([:SOURce])
Chapter 4284
:POWer:REFerence
Supported All
[:SOURce]:POWer:REFerence <val><unit>
[:SOURce]:POWer:REFerence?
This command sets the current output power reference.
*RST
+
0.00000000E
+
000
Range
400 to 300DBM
Key Entry
Ampl Ref Set
Remarks The power reference range is affected by power offset.
:POWer:REFerence:STATe
Supported All
[:SOURce]:POWer:REFerence:STATe ON|OFF|1|0
[:SOURce]:POWer:REFerence:STATe?
This command enables or disables the RF output reference.
*RST 0
Key Entry
Ampl Ref Off On
Remarks Once the reference state is ON, all subsequent output power settings
are set relative to the reference value.
Amplitude offsets can be used with the amplitude reference mode.
Choices ON OFF 1 0