Command Reference
Frequency Subsystem ([:SOURce])
Chapter 4238
:FREQuency:MULTiplier
Supported All
[:SOURce]:FREQuency:MULTiplier <val>
[:SOURce]:FREQuency:MULTiplier?
This command sets the multiplier for the signal generator’s carrier frequency.
*RST
+
1
Key Entry
Freq Multiplier
Remarks For any multiplier other than one, the MULT indicator is shown in the
frequency area of the display.
:FREQuency:OFFSet
Supported All
[:SOURce]:FREQuency:OFFSet <val><unit>
[:SOURce]:FREQuency:OFFSet?
This command sets the frequency offset.
*RST
+
0.0000000000000E
+
00
Key Entry
Freq Offset
Remarks A frequency offset can be entered at any time during normal operation
and when you are operating in frequency reference mode.
When an offset has been entered, the OFFS indicator is turned on in the
frequency area of the display.
The frequency offset state is turned on when any non-zero value is
entered; entering zero will turn it off. Refer to
“:FREQuency:OFFSet:STATe” on page 239 for setting the offset state
independent of entering offset values.
Range Negative Values: –1000 to –.001 Positive Values: .001–1000
Range 20 GHz Models: 0HZ–20GHZ 40 GHz Models: 0HZ–40GHZ
Kommentare zu diesen Handbüchern