Agilent 1100 Series Valves 25
Maintenance, Repair, and Troubleshooting 3
Maintenance and Repair
Reassembling the valve head
1 Mount the new Rotor Seal (5) with the grooves facing the Stator Head (2).
The three pins on the Shaft Assembly fit into the matching holes in the
Rotor Seal only one way.
2 Mount the Stator Ring on the Shaft Assembly that the two pins fit to the
matching holes on the Valve body.
3 Assemble the Stator Head (2) that the pin in the Stator Ring fits to the
matching hole in the Stator Head.
4 Insert the Stator Screws (1). Turn each of the screws an equal amount until
the they are finger-tight, then tighten them another half turn.
Tabl e 6 Rebuild kits and repair parts for Agilent 1100 Series Vales
Description part number
Rebuilt kits
and rotor seals
Rebuild kit for G1157 (2 position/10 port valve) 0101-1359
Rebuild kit for G1158 (2 position/6 port valve) 0101-1358
Rebuild kit for G1159 (6 position selection valve) 0101-1290
Rebuild kit for G1160 (12 position/13 port valve) 0101-1288
Rotor seals for G1158 (2 position/6 port valve) 0100-1855 (Vespel)
0100-1854 (Tefzel)
0100-2233 (PEEK)
Rotor seal for G1162A (2 position/6 port micro
valve)
0100-2087
Rotor seal for G1163A (2 position/10 port micro
valve)
0101-1361