Agilent 1100 Series Valves 23
Maintenance, Repair, and Troubleshooting 3
Maintenance and Repair
2 Align the Stator Ring (4) that the two short pins on the ring enter the
matching holes in the body (7).
3 Insert the three Stator Ring Screws (3). Turn each of the screws an equal
amount until the they are finger-tight, then tighten them another half turn.
4 Mount the new Stator Face (5) onto the Stator Head (2). The pins on the
Stator face must fit into the matching holes of the Stator Head.
5 Assemble the Stator Head (2) and Stator Face(5) that the pin in the Stator
Ring enters the matching hole in the Stator Head.
6 Insert the Stator Screws (1). Turn each of the screws an equal amount until
the they are finger-tight, then tighten them another half turn.
Figure 11 Valve body (7) and Rotor seal (6)
8
6
9
7