38 Chapter 3
Source List Sequencer Parameter Definitions
Source Parameters
Source List Sequencer Parameter
Definitions
3: Transition Time (also known as Setup Time)
This parameter specifies the time required within a given step to allow the source to settle at the specified
frequency and amplitude settings. Recommended settings are as follows:
Frequency change: 0.5 ms
Amplitude change to within 1.0 dB: 0.1 ms
Amplitude change to within 0.1 dB: 0.02 ms
NOTE Although different time units can be used in setting the transition time from the front
panel, the parameter value as entered in the sequence is always in ms.
The key path for selecting Setup Time from the front panel menus is:
[Source], List Sequencer, List Sequencer Setup, Transition Time