36 Chapter 3
Source List Sequencer Parameter Definitions
Source Sequence
Source List Sequencer Parameter
Definitions
Source Sequence
A source sequence consists of a series of steps, during each of which an RF waveform is generated by
the test set’s source.
The steps in a source sequence make it possible to generate a succession of RF waveforms, at a variety of
frequencies and power levels, either independently of the analyzer or in coordination with the data
acquisitions which are controlled by the analyzer list sequencer.
The source can generate an output to the device under test at the same time that the analyzer is receiving
data from it; this makes it possible to test a mobile device in simultaneous transmit and receive mode.