Chapter 11 151
Example Test Scenarios
Calibration Example
Example Test Scenarios
Figure 11-2. The First 2 of 16 Channel Outputs from DUT
Once the DUT has been commanded to begin sending out the uplink signals, we can use the Sequence
Studio application to create a sequence. We begin by creating a single acquisition, with the frequency set
to 824.7 MHz (the uplink frequency for the first channel), and the acquisition set up to trigger on the RF
burst from the DUT. By downloading this simple sequence to the test set and reading back the sequence
trace from the test set, we obtain a graphical display of the uplink signal and its 5 descending levels.
Figure 11-3. The First of 16 Acquisitions in the Sequence