Obsolete and Discontinued Commands 33
Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide 711
:CHANnel2:SKEW
(see page 790)
Command Syntax
:CHANnel2:SKEW <skew value>
<skew value> ::= skew time in NR3 format
<skew value> ::= -100 ns to +100 ns
The :CHANnel2:SKEW command sets the skew between channels 1 and 2.
The maximum skew is +/- 100 ns. You can use the oscilloscope's analog
probe skew control to remove cable delay errors between channel 1 and
channel 2.
Query Syntax
:CHANnel2:SKEW?
The :CHANnel2:SKEW? query returns the current probe skew setting for
the selected channel.
Return Format
<skew value><NL>
<skew value> ::= skew value in NR3 format
See Also "Introduction to :CHANnel<n> Commands" on page 218
NOTE
The :CHANnel2:SKEW command is an obsolete command provided for compatibility to
previous oscilloscopes. Use the :CHANnel<n>:PROBe:SKEW command (see page 230)
instead.
NOTE
This command is only valid for the two channel oscilloscope models.