:SBUS<n> Commands 25
Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide 483
:SBUS<n>:LIN:SIGNal:BAUDrate
(see page 790)
Command Syntax
:SBUS<n>:LIN:SIGNal:BAUDrate <baudrate>
<baudrate> ::= integer from 2400 to 625000 in 100 b/s increments
The :SBUS<n>:LIN:SIGNal:BAUDrate command sets the standard baud rate
of the LIN signal from 2400 b/s to 625 kb/s in 100 b/s increments. If you
enter a baud rate that is not divisible by 100 b/s, the baud rate is set to
the nearest baud rate divisible by 100 b/s.
Query Syntax
:SBUS<n>:LIN:SIGNal:BAUDrate?
The :SBUS<n>:LIN:SIGNal:BAUDrate? query returns the current LIN baud
rate setting.
Return Format
<baudrate><NL>
<baudrate> ::= integer from 2400 to 625000 in 100 b/s increments
See Also "Introduction to :TRIGger Commands" on page 589
":TRIGger:MODE" on page 598
":SBUS<n>:LIN:TRIGger" on page 487
":SBUS<n>:LIN:SOURce" on page 484