Agilent Technologies 2000 X-Series Wartungshandbuch Seite 593

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 930
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 592
:TRIGger Commands 29
Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide 593
:TRIGger:HFReject
(see page 790)
Command Syntax
:TRIGger:HFReject <value>
<value> ::= {{0 | OFF} | {1 | ON}}
The :TRIGger:HFReject command turns the high frequency reject filter off
and on. The high frequency reject filter adds a 50 kHz low- pass filter in
the trigger path to remove high frequency components from the trigger
waveform. Use this filter to remove high- frequency noise, such as AM or
FM broadcast stations, from the trigger path.
Query Syntax
:TRIGger:HFReject?
The :TRIGger:HFReject? query returns the current high frequency reject
filter mode.
Return Format
<value><NL>
<value> ::= {0 | 1}
See Also "Introduction to :TRIGger Commands" on page 589
":TRIGger[:EDGE]:REJect" on page 604
Seitenansicht 592
1 2 ... 588 589 590 591 592 593 594 595 596 597 598 ... 929 930

Kommentare zu diesen Handbüchern

Keine Kommentare