Agilent 75000 Series B
Agilent E1351A/52A/53A/57A/58A
FET Multiplexers
Service Manual
Enclosed is the Service Manual for the Agilent E1351A/52A/53A/57A/58A
FET Multiplexers. Insert this manual, along with any other VXIbus manuals
that you have, into the binder that came with your Agilent Mainframe.
*E1351-90011*
Manual Part Number: E1351-90011
Printed in Malaysia E0606
Seitenansicht 0
1 2 3 4 5 6 ... 71 72

Inhaltsverzeichnis

Seite 1 - *E1351-90011*

Agilent 75000 Series BAgilent E1351A/52A/53A/57A/58AFET MultiplexersService Manual Enclosed is the Service Manual for the Agilent E1351A/

Seite 3

Agilent 75000 Series B Service DocumentationSuggested Sequence to Use ManualsManual DescriptionsInstallation and Getting Started Guide. This manual

Seite 4 - Safety Symbols

What’s in this ManualManual Overview This manual shows how to service the Agilent E1351A, E1352A, E1353A, E1357A, and E1358A FETMultiplexers. Consult

Seite 5 - DECLARATION OF CONFORMITY

Table of ContentsChapter 1 — General InformationIntroduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Seite 6

Chapter 3 — Replaceable partsIntroduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Seite 7

Chapter 1General InformationIntroduction This manual contains information required to test, troubleshoot, and repairthe Agilent E1351A, E1352A, E1353A

Seite 8

SafetyConsiderationsThis product is a Safety Class I instrument that is provided with a protectiveearth terminal when installed in the mainframe. Chec

Seite 9

WARNINGUSING AUTOTRANSFORMERS. If the mainframe is to beenergized via an autotransformer (for voltage reduction) makesure the common terminal is conne

Seite 10

FET Multiplexer DescriptionThe Agilent E1351A, E1352A, E1353A, E1357A, and E1358A FETMultiplexer is an "instrument" in a VXIbus mainframe. A

Seite 11 - Manual Descriptions

Agilent E1357ADescriptionThe Agilent E1357A provides up to eight channels of strain gage switching.Strain gage measurements are supported in 1/4 bridg

Seite 13 - Table of Contents

RecommendedTest EquipmentTable 1-1 lists the test equipment recommended for testing, adjusting, andservicing the FET Multiplexers. Essential requireme

Seite 14 - Chapter 4 — Service

Notify AgilentNotify Agilent and carrier.Figure 1-2. Initial (Incoming) Inspection GuidelinesGeneral Information 1-7

Seite 15 - General Information

ShippingGuidelinesFollow the steps in Figure 1-3 to return a FET Multiplexer to an AgilentTechnologies Sales and Support Office or Service Center. 1

Seite 16 - Considerations

Chapter 2Verification TestsIntroduction This chapter describes the verification tests for the Agilent E1351A,E1352A, E1353A, E1357A, and E1358A. The t

Seite 17 - General Information 1-3

FunctionalVerification TestThe Functional Verification Test for the Agilent E1351A, E1352A,E1353A, E1357A, and E1358A FET Multiplexers consists of sen

Seite 18 - Description

OperationVerification TestThe procedures in this section are used to provide a high level of confidencethat the FET Multiplexer is meeting published s

Seite 19 - General Information 1-5

re-wire each time the tests are performed. The Agilent E1351A terminalblock and case assembly part number is E1351-80001.Test 2-1: ClosedChannelResist

Seite 20 - Shipping

• Repeat step 3 for channels 01 through 15 • Use CLOS (@nncc), where nn is the card number and cc is thechannel numberLO ChannelMeasurements 1. Make

Seite 21 - General Information 1-7

Tree Switch HI ChannelMeasurements 1. Make Hardware Connections• Turn mainframe power OFF• Connect DMM as shown in Figure 2-4• Turn mainframe power O

Seite 22 - Guidelines

Tree Switch LOChannel Measurements 1. Make Hardware Connections• Turn mainframe power OFF• Connect DMM as shown in Figure 2-5• Turn mainframe power O

Seite 23 - Verification Tests

CertificationAgilent Technologies certifies that this product met its published specifications at the time of shipment from the factory. AgilentTechn

Seite 24 - Verification

Example: ClosedChannel ResistanceTestThis example performs a closed channel resistance test of all measurementpaths. If a FET on resistance is >3.1

Seite 25 - Verification Tests 2-3

370 OUTPUT @Dmm;"TRIG SGL"380 ENTER @Dmm;Value390 IF Value<10000 THEN400 CLEAR SCREEN410 PRINT "Measurement indi

Seite 26 - Resistance Test

780 PAUSE790 OUTPUT @Dmm;"PRESET NORM;FUNC OHM"800 OUTPUT @Mux;"*RST"810 !820 ! Check for stuck tree switches8

Seite 27 - Verification Tests 2-5

Typical ResultClosed channel resistance measurement resultsChannel 0 HI 989 Ohms LO 1004 OhmsChannel 1 HI 991 Ohms LO 979 OhmsChannel 2 HI 1001 Ohms L

Seite 28 - 2-6 Verification Tests

Test 2-2: LeakageTestThe test verifies the input impedance by measuring the voltage drop across aknown resistor in series with the input impedance. Le

Seite 29 - Verification Tests 2-7

3. Check Channels Leakage • Send CLOS(@nn00) to the FET Multiplexer to close channel00, where nn is card number (typically 01)• Send TRIG SGL to DMM•

Seite 30 - 2-8 Verification Tests

HI to Chassis Leakage 1. Make Hardware Connections• Turn power supply and mainframe power OFF• Connect DMM, power supply, and resistor as shown in F

Seite 31 - Verification Tests 2-9

3. Check Channels Leakage • Send CLOS(@nn00) to the FET Multiplexer, where nn is thechannel number (typically 01)• Send TRIG SGL to DMM• Record the D

Seite 32 - 2-10 Verification Tests

• Send CLOS(@nn15) to the FET Multiplexer to close channel15 and tree B• Send TRIG SGL to DMM• Record the DMM reading in Table 2-1 Positive Polarity,

Seite 33 - Verification Tests 2-11

LO to Chassis Leakage 1. Make Hardware Connections• Turn power supply and mainframe power OFF• Connect DMM, power supply, and resistor as shown in

Seite 34 - Test 2-2: Leakage

Frame or chassis ground terminal—typicallyconnects to the equipment’s metal frame.Alternating current (AC).Direct current (DC). Indicates hazardous vo

Seite 35 - Verification Tests 2-13

• Record the DMM reading in Table 2-1 Positive Polarity, LO to Chassis, Channels• Send OPEN (@nn00) to the FET Multiplexer4. Check Tree Leakage Curre

Seite 36 - 2-14 Verification Tests

Example: LeakageCurrent TestThis example performs a leakage test from HI to LO, HI to Chassis, and LOto Chassis. If the leakage is too high (caused by

Seite 37 - Verification Tests 2-15

170 PRINT " 1. Turn mainframe, power supply, and DMM power OFF" 180 PRINT " 2. Connect GPIB cable between mainframe and DMM

Seite 38 - 2-16 Verification Tests

580 END IF 590 OUTPUT @Mux;"OPEN (@"&Cc$&"00)" 600 OUTPUT @Mux;"CLOS (@"&Cc$&"15)" 61

Seite 39 - Verification Tests 2-17

Typical ResultPositive polarity leakageDirect Channels Tree A Tree BDIRECT HI to DIRECT LO 0.0021 Vdc 0.0015 Vdc 0.0020 Vdc 0.0018 VdcHI to CHASSIS 0.

Seite 40 - 2-18 Verification Tests

Leakage Test Conditions:• DC Volts function, 100 mV range• 90 day specifications• Worst case reading = 0.01 V• Resistor value: 99 kΩ < R < 101

Seite 41 - Verification Tests 2-19

Table 2-1. Performance Test Record (Page 1 of 2)General InformationTest Facility:Name _____________________________________Address __________________

Seite 42 - 2-20 Verification Tests

Table 2-1. Performance Test Record (Page 2 of 3)Test No/Description Minimum*ValueMeasured Value (V) MaximumValue MeasUncert Test AccRatio (TAR)2-1.

Seite 43 - Verification Tests 2-21

Table 2-1. Performance Test Record (Page 3 of 3)Test No/Description MinimumValue *Measured Value (V) MaximumValue MeasUncert Test AccRatio (TAR)2-2

Seite 44 - Test Record

Chapter 3Replaceable PartsIntroduction This chapter contains information to order replaceable parts for the AgilentE1351A, E1352A, E1353A, E1357A, and

Seite 45 - Ratio (TAR)

DECLARATION OF CONFORMITYAccording to ISO/IEC Guide 22 and CEN/CENELEC EN 45014

Seite 46 - 2-24 Verification Tests

Table 3-1. FET Multiplexer Replaceable Parts ReferenceDesignatorAgilent PartNumberQty Part Description Mfr.CodeMfr. PartNumberAgilent E1351AA1 E1351-

Seite 47 - Verification Tests 2-25

ReferenceDesignatorAgilent PartNumberQty Part Description Mfr.CodeMfr. PartNumberE1351-66201 1 16-CHANNEL FET MULTIPLEXER(See figure 3-1)28480 E1351-6

Seite 48 - 2-26 Verification Tests

Table 3-4. FET Multiplexers Reference Designators FET Multiplexers Reference DesignatorsA... assembly

Seite 49 - Replaceable Parts

Mechanical Parts LocatorsFigure 3-1 shows the mechanical replaceable parts on the componentassembly (Agilent PN E1351-66201). Figure 3-2 shows the mec

Seite 50 - 3-2 Replaceable Parts

Figure 3-2. Terminal Case Assembly Replaceable Parts3-6 Replaceable Parts

Seite 51 - Replaceable Parts 3-3

Chapter 4ServiceIntroduction This chapter contains service information for the Agilent E1351A, E1352A,E1353A, E1357A, and E1358A FET Multiplexers. Als

Seite 52 - 3-4 Replaceable Parts

Figure 4-1. Agilent E1351A Simplified Switch Diagram4-2 Service

Seite 53 - Locators

Figure 4-2. Agilent E1352A Simplified Switch DiagramService 4-3

Seite 54 - 3-6 Replaceable Parts

Figure 4-3. Agilent E1353A Simplified Switch Diagram4-4 Service

Seite 55 - Chapter 4

Figure 4-4. Agilent E1357A/58A Simplified Switch DiagramService 4-5

Seite 56 - 4-2 Service

DECLARATION OF CONFORMITYAccording to ISO/IEC Guide 22 and CEN/CENELEC EN 45014

Seite 57 - Service 4-3

for use with the Agilent E1357A and Agilent E1358A Strain Gage FETMultiplexers. Repair Strategy Agilent Technologies recommends replacement of the en

Seite 58 - 4-4 Service

Making VisualChecksVisual checks for the FET Multiplexers include the following. See Table 4-2 for typical checks.• Check switches/jumpers• Check for

Seite 59 - Service 4-5

Repair and Maintenance GuidelinesThis section provides guidelines for repairing and maintaining the FETMultiplexer including:• ESD precautions• Solder

Seite 60 - Techniques

Chapter AVerification Tests - C ProgramsFunctionalVerificationTestThis program is designed to do the Functional Verification Test found inChapter 2 -

Seite 61 - Making Visual

PerformanceVerificationTestsThese programs are designed to do the Performance Verification Testsfound in Chapter 2 - Verification Tests.NOTEThese prog

Seite 62 - Maintenance

/*...Measure Closed Channel Resistance...*/ iprintf (id, "*RST\n"); iprintf (dm, "PRESET NORM;TRIG

Seite 63 - Chapter A

printf ("\n\nMeasurements complete for Channel HI and LO"); printf ("\nPress ENTER for Tree Switch measurements"); getchar ();

Seite 64

if (tree[i][j] > 3100) { if (j == 0) printf ("\n*** Resistance for A Tree Switch is 3.1 kOhms"); else

Seite 65 - Example C Programs A-3

Example: LeakageCurrent TestThis example performs a leakage test from HI to LO, HI to Chassis, and LOto Chassis. If the leakage is too high (caused b

Seite 66 - A-4 Example C Programs

if (i == 0) { printf ("\n2. Connect DMM input HI leads to Power Supply HI"); printf ("\n3. Connect DMM input LO leads

Seite 67 - Example C Programs A-5

DECLARATION OF CONFORMITYAccording to ISO/IEC Guide 22 and CEN/CENELEC EN 45014

Seite 68 - Example: Leakage

if (result[i][1] > .01) printf ("\n\n***** Channel path leakage out of tolerance %6.4lf Volts *****",result[i][1]); /*...

Seite 69 - Example C Programs A-7

printf ("\n\n-------------------------------------------------------------"); printf ("\n Negative Polarity Leaka

Seite 70 - A-8 Example C Programs

A-10 Example C Programs

Seite 71 - Example C Programs A-9

DECLARATION OF CONFORMITYAccording to ISO/IEC Guide 22 and CEN/CENELEC EN 45014

Seite 72 - A-10 Example C Programs

DECLARATION OF CONFORMITYAccording to ISO/IEC Guide 22 and CEN/CENELEC EN 45014 viiManufacturer’s Name:Agilent Te