:EXTernal Trigger Commands 15
Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide 273
:EXTernal:PROBe
(see page 790)
Command Syntax
:EXTernal:PROBe <attenuation>
<attenuation> ::= probe attenuation ratio in NR3 format
The :EXTernal:PROBe command specifies the probe attenuation factor for
the external trigger. The probe attenuation factor may be 0.1 to 10000.
This command does not change the actual input sensitivity of the
oscilloscope. It changes the reference constants for scaling the display
factors and for setting trigger levels.
If an AutoProbe probe is connected to the oscilloscope, the attenuation
value cannot be changed from the sensed value. Attempting to set the
oscilloscope to an attenuation value other than the sensed value produces
an error.
Query Syntax
:EXTernal:PROBe?
The :EXTernal:PROBe? query returns the current probe attenuation factor
for the external trigger.
Return Format
<attenuation><NL>
<attenuation> ::= probe attenuation ratio in NR3 format
See Also "Introduction to :EXTernal Trigger Commands" on page 271
":EXTernal:RANGe" on page 274
"Introduction to :TRIGger Commands" on page 589
":CHANnel<n>:PROBe" on page 227