Agilent Technologies 2000 X-Series Wartungshandbuch Seite 326

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 930
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 325
326 Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide
19 :MARKer Commands
:MARKer:XUNits:USE
(see page 790)
Command Syntax
:MARKer:XUNits:USE
When DEGRees is selected for :MARKer:XUNits, the :MARKer:XUNits:USE
command sets the current X1 location as 0 degrees and the current X2
location as 360 degrees.
When PERCent is selected for :MARKer:XUNits, the :MARKer:XUNits:USE
command sets the current X1 location as 0 percent and the current X2
location as 100 percent.
Once the 0 and 360 degree or 0 and 100 percent locations are set, inputs
to and outputs from the :MARKer:X1Position, :MARKer:X2Position, and
:MARKer:XDELta commands/queries are relative to the set locations.
See Also "Introduction to :MARKer Commands" on page 318
":MARKer:XUNits" on page 325
":MARKer:X1Y1source" on page 321
":MARKer:X2Y2source" on page 323
":MEASure:SOURce" on page 361
":MARKer:X1Position" on page 320
":MARKer:X2Position" on page 322
":MARKer:XDELta" on page 324
Seitenansicht 325
1 2 ... 321 322 323 324 325 326 327 328 329 330 331 ... 929 930

Kommentare zu diesen Handbüchern

Keine Kommentare