C Specifications Definition of Terms / Measurement Conditions
192 E6000C Mini-OTDR Users Guide, E0302
Definition of Terms /
Measurement Conditions
Generally, the wavelengths are given by the specific
module. Therefore, the measurement conditions do not
contain the wavelength. Unless otherwise limited, all
specifications are valid for the specified environmental
conditions.
All data presented in the ± form are to be understood as
peak-to-peak variation divided by 2.
Attenuation deadzone
The distance from the start of a reflective event to the
point where the receiver has recovered to within a ±0.5
dB margin around the undisturbed and averaged
backscatter trace.
Conditions: Reflective, lossless event with specified
reflectance, at specified instrument settings.
±0.5 dB margin
around averaged
backscatter trace
Attenuation
Deadzone
Distance