Chapter 1 Performance Tests
Agilent 53210A/53220A/53230A Complete Performance Tests
Assembly-Level Service Guide 71
Test 7c: (Q) Pulse Burst Frequency - 53230A Channel 3
Option 150 With Option 115
This test uses the 53230A to measure both narrow and wide pulse outputs
of a 1.35 GHz carrier using the Opt 115 microwave channel.
NOTE This test procedure applies only to the 53230A Counter with
Option 115 and Option 150 installed. Option 115 is installed as
microwave channel 3.
NOTE The Quick Verification (Q) requires only the Narrow Pulse Width
Burst Frequency Test be performed.
Equipment
• FS725 Rubidium Timebase
• Agilent E8257D-520-UNW Microwave Signal Generator
• Agilent 10503A BNC Cable (2)
• Agilent 11500E (3.5) mm cable
• Agilent 1250-1250 N(M) to SMA(F)
E8257D-520-UNW Setup
• Output Frequency: 1.35 GHz
•Output Power: -6 dBm
• Auto-Level Control: Off
•Pulse Mode: On
• Pulse Repetition Interval Period: 1 mSec
• Pulse Width (Narrow): 400 nSec