
Chapter 1 Performance Tests
Agilent 53210A/53220A/53230A Complete Performance Tests
28 Assembly-Level Service Guide
7 Press and set:
• Noise Rejection: Off
8Press the Frequency/Period key.
• The following softkeys are displayed:
•If Advanced is appears, press Auto to select Auto measurement mode.
NOTE The 53210A does not have an Auto measurement mode.
9 Press the Channel 1 input key.
• The following softkeys are displayed:
10 Press and set:
• Coupling: DC
• Input Impedance: 50
• Range: 5 V
• Level Setup / Auto Level: Off
• Level Setup / Input Level: 0.000 V
•BW Limit: Off
•Probe: None
Kommentare zu diesen Handbüchern