Agilent Technologies 4294A Bedienerhandbuch Seite 153

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 469
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 152
Chapter 6 145
Calibration
G. Calibration When the 3.5-mm Port of the 42941A is Extended
6. Calibration
G. Calibration When the 3.5-mm Port of the 42941A is
Extended
Perform the following calibration when the 3.5-mm port of a 42941A connected to the
Agilent 4294A is extended with a one-port configuration and the device is connected to the
cable’s end with a minimal connection.
Figure 6-7 Calibration When the 3.5-mm Port of the 42941A is Extended
Step 1. Make sure that the adapter selection is set to “42941A Impedance Probe (PROBE
42941A).”
See “Adapter Setting” on page 67 for adapter selection.
Seitenansicht 152
1 2 ... 148 149 150 151 152 153 154 155 156 157 158 ... 468 469

Kommentare zu diesen Handbüchern

Keine Kommentare