Agilent Technologies 4294A Spezifikationen

Stöbern Sie online oder laden Sie Spezifikationen nach Multimeter Agilent Technologies 4294A herunter. Agilent Technologies 4294A Specifications Benutzerhandbuch

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken

Inhaltsverzeichnis

Seite 1 - Handbook

Agilent TechnologiesImpedance MeasurementHandbookJuly 2006

Seite 2

1-4. True, effective, and indicated valuesA thorough understanding of true, effective, and indicated values of a component, as well as theirsignifican

Seite 3

Figure 5-36. Measurement setup5-26

Seite 4

5-11. Battery measurementThe internal resistance of a battery is generally measured using a 1 KHz AC signal. When a batteryis connected directly to t

Seite 5

5-12. Test signal voltage enhancementWhen measuring the impedance of test signal level dependent devices, such as liquid crystals, induc-tors and high

Seite 6

Figure 5-39. Connection diagram of test signal voltage enhancement circuit5-29

Seite 7 - Impedance measurement basics

5-13. DC bias voltage enhancementDC biased impedance measurement is popularly used to evaluate the characteristics of the deviceunder the conditions w

Seite 8

Figure 5-40. External DC bias measurement setupExternal DC voltage bias protection in 4TP configurationIf the measurement frequency is above 2 MHz or

Seite 9 - 1-2. Measuring impedance

5-14. DC bias current enhancementDC current biasing is used for inductor and transformer measurement. In low frequency region, theE4980A or 4284A pre

Seite 10

Take caution of electrical shock hazards when using the external DC bias circuit.A large energy is charged in L1 and L2 as well as the DUT (Lx) by a b

Seite 11

5-15. Equivalent circuit analysis and its applicationAgilent’s impedance analyzers are equipped with an equivalent circuit analysis function. The pur

Seite 12

Figure 5-44. Equivalent circuit modelsIf the simulated frequency response curve partially fits the measurement results, it can be said thatthe selecte

Seite 13

1-5. Component dependency factorsThe measured impedance value of a component depends on several measurement conditions, suchas frequency, test signal

Seite 14

Figure 5-45. Frequency response simulation for a low-value inductorFigure 5-46. Equivalent circuit enhancement5-3610 DIM Ztrc(1:201,1:2),Fmta$[9],Fmtb

Seite 15 - SECTION 2

Measurement accuracy can be improved by taking advantage of the equivalent circuit analysis.Figure 5-47 (a) shows an Ls-Q measurement example for an i

Seite 16 - Network analysis method

This page intentionally left blank.

Seite 17 - Auto balancing bridge method

APPENDIX AThe concept of a test fixture’s additional error1. System configuration for impedance measurement Very often, the system configured for impe

Seite 18

A-2The equation for the test fixture’s additional error is shown below:Ze = ± { A + (Zs/Zx + Yo•Zx) × 100} (%)De = Ze/100 (D ≤ 0.1)Ze : Additional er

Seite 19

A-3Open offset error:The term, Yo•Zx × 100 is called open offset error. If the same analysis is carried out with admittance,then it can be concluded

Seite 20

A-4Terminal connection method:In order to make short repeatability small, there are test fixtures which utilize the 4-terminal con-nection method (for

Seite 21

A-5obtained. For open repeatability, measure the admittance of the test fixture’s open condition. In thesame way, determine open repeatability by meas

Seite 22

This page intentionally left blank.

Seite 23

B-1APPENDIX BOpen and short compensationThe open/short compensation used in Agilent’s instrument models the residuals as a linear networkrepresented b

Seite 24

Figure 1-9. Capacitor frequency responseTest signal level:The test signal (AC) applied may affect the measurement result for some components. For exa

Seite 25

B-2Figure B-1. Open/short compensation (2 of 2)

Seite 26

C-1APPENDIX COpen, short and load compensationThe open/short/load compensation requires the measurement data of a standard DUT with knownvalues in add

Seite 27

This page intentionally left blank.

Seite 28

D-1APPENDIX DElectrical length compensationA test port extension can be modeled using a coaxial transmission line as shown in Figure D-1.When an imped

Seite 29

D-2When a (virtual) transmission line in which the signal wavelength is equal to the wavelength in vac-uum is assumed, the virtual line length ( e) th

Seite 30

E-1APPENDIX EQ Measurement accuracy calculationQ measurement accuracy for auto balancing bridge type instruments is not specified directly as ±%.Q

Seite 31

www.agilent.com/find/emailupdatesGet the latest information on the products andapplications you select. Agilent Email Updateswww.agilent.comAgilent T

Seite 32

Figure 1-11. DC bias dependencies of ceramic capacitors and cored-inductorsTemperature:Most types of components are temperature dependent. The temper

Seite 33

This page intentionally left blank.

Seite 34

SECTION 2Impedance measurement instruments2-1. Measurement methodsThere are many measurement methods to choose from when measuring impedance, each of

Seite 35

2-2While the RF I-V measurement method is based on the sameprinciple as the I-V method, it is configured in a differentway by using an impedance match

Seite 36

Figure 2-1. Impedance measurement method (3 of 3)2-3The current, flowing through the DUT, also flows through resistor R. Thepotential at the “L” poin

Seite 37 - Fixturing and cabling

Table 2-1. Common impedance measurement methodsNote: Agilent Technologies currently offers no instruments for the bridge method and the resonant metho

Seite 38

2-2. Operating theory of practical instrumentsThe operating theory and key functions of the auto balancing bridge instrument are discussed in theparag

Seite 39

This page intentionally left blank.

Seite 40

The measurement circuit is functionally divided into following three sections.The signal source section generates the test signal applied to the unkno

Seite 41 - 3-3. Test fixtures

Figure 2-4. Auto balancing bridge section block diagramFigure 2-5. Vector ratio detector section block diagram2-7

Seite 42

2-4. Key measurement functionsThe following discussion describes the key measurement functions for advanced impedance mea-surement instruments. Thoro

Seite 43

2-4-2. DC biasIn addition to the AC test signal, a DC voltage can be output through the Hc terminal and applied tothe DUT. A simplified output circui

Seite 44 - 3-4. Test cables

2-4-3. Ranging functionTo measure impedance from low values to high values, impedance measurement instruments haveseveral measurement ranges. General

Seite 45

2-4-4. Level monitor functionMonitoring the test signal voltage or current applied to the DUT is important for maintaining accu-rate test conditions,

Seite 46 - Figure 3-9. 4TP-4TP extension

Averaging function calculates the mean value of measured parameters from the desired number ofmeasurements. Averaging has the same effect on random n

Seite 47

The induced errors are dependent upon test frequency, test fixture, test leads, DUT connection con-figuration, and surrounding conditions of the DUT.

Seite 48

Figure 2-11. Guarding techniques2-4-8. Grounded device measurement capabilityGrounded devices such as the input/output of an amplifier can be measured

Seite 49 - 3-7. RF test fixtures

Figure 2-12. Low-grounded device measurement2-15

Seite 50

iThe Impedance Measurement HandbookA Guide to Measurement Technology and TechniquesCopyright®2000-2003 Agilent Technologies Co. LtdAll rights reserve

Seite 51

2-5. Theory of RF I-V measurement methodThe RF I-V method featuring Agilent’s RF impedance analyzers and RF LCR meters is an advancedtechnique to meas

Seite 52

Figure 2-13. Simplified block diagram for RF I-V method2-17

Seite 53 - SECTION 4

2-6. Difference between RF I-V and network analysis measurement methodsWhen testing components in the RF region, the RF I-V measurement method is ofte

Seite 54

Figure 2-14. Relationship of reflection coefficient to impedanceFigure 2-15. Measurement sensitivity of network analysis and RF I-V methods2-19

Seite 55

Figure 2-16. Comparison of typical Q accuracy2-7. Key measurement functions2-7-1. OSC levelThe oscillator output signal is output through the coaxial

Seite 56

2-7-3. CalibrationMost of the RF vector measurement instruments such as network analyzers need to be calibratedeach time a measurement is initiated or

Seite 57

This page intentionally left blank.

Seite 58

SECTION 3Fixturing and cablingWhen interconnecting a device under test (DUT) to the measurement terminals of the auto balancingbridge instrument, ther

Seite 59

The four-terminal pair (4TP) configuration solves the mutual coupling problem because it uses coaxialcable to isolate the voltage sensing cables from

Seite 60

Figure 3-2. Three-terminal (3T) configurationFigure 3-3. Four-terminal (4T) configuration3-3

Seite 61

iiSECTION 3 Fixturing and cabling––––––– LF impedance measurement –––––––Paragraph 3-1 Terminal configuration ...

Seite 62

Figure 3-4. Five-terminal (5T) configurationFigure 3-5. Four-terminal pair (4TP) configuration3-4

Seite 63

3-2. Using test cables at high frequenciesThe 4TP configuration is the best solution for wide-range impedance measurement. However, inbasic 4TP measu

Seite 64

Table 3-1. Test fixture’s DUT connection configuration and applications3-3-2. User fabricated test fixturesIf the DUT is not applicable to Agilent sup

Seite 65

3-3-3. User test fixture exampleFigure 3-7 shows an example of a user fabricated test fixture. It is equipped with alligator clips asthe contact elec

Seite 66

3-4. Test cablesWhen the DUT is tested apart from the instrument, it is necessary to extend the test ports(UNKNOWN terminals) using cables. If the ca

Seite 67

Figure 3-8. Specifications of recommended cable (Agilent PN 8121-1218)3-4-3. Test cable extensionIf the required test cable is longer than 1 m, 2 m, o

Seite 68

Figure 3-9. 4TP-4TP extensionFigure 3-10. Shielded 2T extension3-10

Seite 69

Figure 3-11. Shielded 4T extensionTable 3-2. Summary of cable extension3-11Measurement frequency100 kHz and below 100 kHz and aboveLowTypically 4TP-4T

Seite 70

3-5. Eliminating the stray capacitance effectsWhen the DUT has high impedance (e.g. Low Capacitance), the effects of stray capacitance are notnegligib

Seite 71

Figure 3-13. Coaxial test port circuit configuration3-7. RF test fixturesRF test fixtures are designed so that the lead length (electrical path length

Seite 72

iii4-7-1 Variance in residual parameter value ... 4-194-7-2 A difference in contact condition ...

Seite 73

3-7-1. Agilent supplied RF test fixturesAgilent Technologies offers various types of RF test fixtures that meet the type of the DUT andrequired test f

Seite 74

3-8. Test port extension in RF regionIn RF measurements, connect the DUT closely to the test port to minimize additional measurementerrors. When ther

Seite 75 - SECTION 5

Figure 3-15. Calibration plane extensionFigure 3-16. Practical calibration and compensation at extended test port3-16

Seite 76 - Table 5-1. Capacitor types

SECTION 4Measurement error and compensation4-1. Measurement errorFor real-world measurements, we have to assume that the measurement result always con

Seite 77

4-2-1. Offset compensationWhen a measurement is affected by only a single component of the residuals, the effective value canbe obtained by simply sub

Seite 78

Figure 4-3. Open/short compensation4-2-3. Precautions for open and short measurementsWhen an open measurement is made, it is important to accurately m

Seite 79 - 5-2. Inductor measurement

Figure 4-4. Example of shorting device. (Agilent PN: 5000-4226)4-2-4. Open, short and load compensationsThere are numerous measurement conditions wher

Seite 80

Figure 4-5. Open/short/load compensation4-2-5. What should be used as the load?The key point in open/short/load compensation is to select a load whose

Seite 81

Figure 4-6. Electrode distance in load measurementFigure 4-7. Load value must be close to DUT’s value4-6

Seite 82

Figure 4-8. Actual open/short load measurement example4-2-6. Application limit for open, short and load compensationsWhen the residuals are too signif

Seite 83 - 5-3. Transformer measurement

ivThis page intentionally left blank.

Seite 84

Figure 4-9. Effect of contact resistance4-8If RH = RL = Rhp = Rlc and Chp = Clp, D errorsof 2-terminal and 4-terminal become the samewhen Cx = ChpThis

Seite 85

4-4. Measurement cable extension induced errorExtending a 4TP measurement cable from the instrument will cause a magnitude error and phaseshift of the

Seite 86

Figure 4-11. Measurement error due to extended cable lengthThe cable length compensation works for test cables whose length and propagation constants

Seite 87 - 5-4. Diode measurement

k value is a decimal number mostly within the range of -1 to +1 and different for different instru-ments. As the above equation shows, the error rapi

Seite 88 - 5-5. MOS FET measurement

Figure 4-12. Compensation examples4-12

Seite 89

4-6. Calibration and compensation in RF region4-6-1. CalibrationWhether the RF I-V method or network analysis, the open, short and load calibration mi

Seite 90

When the test port is extended, calibration should be performed at the end of extension cable, asdiscussed in section 3. Thereby, the calibration pla

Seite 91 - 42941A Impedance Probe

4-6-3. Compensation methodAs the error source model is different for the coaxial and non-coaxial sections of the test fixture,compensation method is a

Seite 92

Accordingly, the residual parameters have greater effects on higher frequency measurements andbecome a primary factor of measurement errors. The accu

Seite 93 - Figure. 5-28

Conceptually, there are two methods of defining the short bar’s impedance: One is to assume theimpedance to be zero. This has been a primordial metho

Seite 94 - 5-8. Resonator measurement

SECTION 1Impedance measurement basics1-1. ImpedanceImpedance is an important parameter used to characterize electronic circuits, components, and thema

Seite 95

4-6-7. Electrical length compensationIn the lower frequency region, using the open/short compensation function can minimize most oftest fixture residu

Seite 96

4-6-8. Practical compensation techniqueThe calibration and compensation methods suitable for measurement are different for how the testcable or fixtur

Seite 97 - 5-9. Cable measurements

Figure 4-17. Difference in residual parameter values due to DUT positioning4-7-2. A difference in contact conditionChange in contact condition of the

Seite 98

4-7-3. A difference in open/short compensation conditionsImproper open/short measurements deteriorate accuracy of compensated measurement results. If

Seite 99

Figure 4-20. Eddy current effect and magnetic flux directivity of device4-7-5. Variance in environmental temperatureTemperature influences on the elec

Seite 100

SECTION 5Impedance measurement applications and enhancementsImpedance measurement instruments are used for a wide variety of applications. In this se

Seite 101 - 5-11. Battery measurement

Table 5-1. Capacitor typesType Application Advantage DisadvantageFilm Blocking, buffering. Wide range of capacitance Medium costbypass, coupling. and

Seite 102

When we measure capacitors, we have to consider these parasitics. Impedance measurementinstruments measure capacitance in either the series mode (Cs-

Seite 103

Precautions for capacitor measurement depend on the capacitance value being measured.High-value capacitance measurement is a low impedance measurement

Seite 104

5-2. Inductor measurementAn inductor consists of wire wound around a core and is characterized by the core material used.Air is the simplest core mate

Seite 105

Reactance takes two forms - inductive (XL) and capacitive (Xc). By definition, XL=2πfL andXc=1/(2πfC), where f is the frequency of interest, L is ind

Seite 106

±0.001, the maximum measurable Q value is 90.9. See Appendix E for Q accuracy calculation equa-tion.) Except for resonant method, the impedance measu

Seite 107

Figure 5-11. Q measurement accuracyFigure 5-12. Q measurement errorFurthermore, the following phenomena may occur when a cored inductor is measured us

Seite 108

Figure 5-13. Harmonic distortion caused by inductor5-8

Seite 109

5-3. Transformer measurementA transformer is one end-product of an inductor. So, the measurement techniques are the same asfor inductor measurement.

Seite 110

Inter-winding capacitance (C) between the primary and the secondary is measured by connecting oneside of each winding to the instrument as shown in Fi

Seite 111

Turns ratio (N) Approximate the turns ratio (N) by connecting a resistor in the secondary as shown inFigure 5-19 (a). From the impedance value measur

Seite 112

The 4263B’s transformer measurement function enables the measurement of the N, M, L1 and theDC resistance of the primary by changing measurement circu

Seite 113 - APPENDIX A

5-4. Diode measurementThe junction capacitance of a switching diode determines its switching speed and is dependent onthe reverse DC voltage applied t

Seite 114

5-5. MOS FET measurementEvaluating the capacitances between the source, drain, and gate of an MOS FET is important indesign of high frequency and swit

Seite 115

5-6. Silicon wafer C-V measurementThe C-V (capacitance vs. DC bias voltage) characteristic of a MOS structure is an important mea-surement parameter f

Seite 116

1-2. Measuring impedanceTo find the impedance, we need to measure at least two values because impedance is a complexquantity. Many modern impedance m

Seite 117 - Type of error Impedance

Figure 5-26. C-V measurement setup5-16

Seite 118

As a result of extremely high integration of logic LSIs using MOS FETs, the thickness of the MOSFETs’ gate oxide is becoming thinner (less than 2.0 nm

Seite 119 - Open and short compensation

5-7. High-frequency impedance measurement using the probeAs shown in Table 5-3, an RF I-V instrument can be used for a wafer’s L, C, and R measurement

Seite 120

Figure. 5-28. 5-19

Seite 121 - APPENDIX C

5-8. Resonator measurementThe resonator is the key component in an oscillator circuit. Crystal and ceramic resonators arecommonly used in the kHz and

Seite 122

2. It is important to properly set the oscillator output level; resonators are test signal dependent.The minimum impedance value and the series resona

Seite 123 - APPENDIX D

Figure 5-31. Resonator equivalent circuit mode5-22(a) (b)

Seite 124

5-9. Cable measurementsThe characteristic impedance Z(Ω), capacitance per unit length C (pF/m) and the propagation con-stants α (dB/m) and β (rad/m) a

Seite 125 - APPENDIX E

Figure 5-33. Measurement resultBalanced cable measurementA balun transformer is required for measuring balanced cable because the instrument’s UNKNOWN

Seite 126 - Agilent Direct

5-10. Balanced device measurementWhen a balanced DUT (such as balanced cable or the balanced input impedance of a differentialamplifier) is measured,

Kommentare zu diesen Handbüchern

Keine Kommentare