Agilent Technologies 4294A Bedienerhandbuch Seite 151

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 469
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 150
Chapter 6 143
Calibration
F. Calibration When a Probe Adapter is Connected to the 42941A
6. Calibration
F. Calibration When a Probe Adapter is Connected to the
42941A
Perform the following calibration when a probe adapter supplied with the 42941A
Impedance Probe is directly connected to the 3.5-mm port of the 42941A, which is
connected to the four-terminal pair port of the Agilent 4294A, and a device is connected to
the probe adapter.
Figure 6-6 Calibration When a Probe Adapter is Connected to the 42941A
This calibration procedure is applied when the following probe adapters are connected to
the 3.5-mm port of the 42941A (see the operation manual of the 42941A for details of
probe adapters):
Pin Probe (with ground electrode or ground lead)
•BNC(m) Adapter
BNC(m) Adapter with alligator clip leads
Step 1. Make sure that the adapter selection is set to “42941A Impedance Probe
(
PROBE 42941A).”
Seitenansicht 150
1 2 ... 146 147 148 149 150 151 152 153 154 155 156 ... 468 469

Kommentare zu diesen Handbüchern

Keine Kommentare