Chapter 3: Testing Performance
Performance Test Record
64
Performance Test Record
Agilent Technologies
Agilent 8000 Series Oscilloscopes
Model Number _____________________ Tested by___________________
Serial Number ___________________________ Work Order No.___________________
Recommended Test Interval - 1 Year/2000 hours Date___________________
Recommended next test date ___________________ Ambient temperature ___________________
Test Limits Results
(Vmax - Vmin)/4.8 =Limits
DC Calibrator
Amplitude
1.000 0.998 to 1.002 ___________________
Channel 1 Channel 2 Channel 3 Chan 4/Ext
Input
Resistance
50
Ω 49.25 Ω to 50.75 Ω ___________ ___________ ___________ ______/_____
1 M
Ω 990 kΩ to 1.010 MΩ ___________ ___________ ___________ ______/_____
Time Scale
Accuracy
Limit
Limit is based on oscilloscope’s age
Calculated
TSA Spec
Measured time
scale error
Pass/Fail
TSA ≤ ± (15 + 2 * YrsSinceMfr) ppm ___________ ___________ ___________
MSO8064A,
DSO8064A
Delta Time
Measurement
Accuracy
Limits Results
Averaged Mode DTMA Channel 1 Channel 2 Channel 3 Channel 4
Maximum (ps) 3326.2 ___________ ___________ ___________ ___________
Minimum (ps) 3318.3 ___________ ___________ ___________ ___________
Std. Deviation (ps rms) 1.2 ___________ ___________ ___________ ___________
Non-Averaged Mode DTMA Channel 1 Channel 2 Channel 3 Channel 4
Maximum (ps) 3352.2 ___________ ___________ ___________ ___________
Minimum (ps) 3292.3 ___________ ___________ ___________ ___________
Std. Deviation 6.8 ___________ ___________ ___________ ___________
MSO8104A,
DSO8104A
Delta Time
Measurement
Accuracy
Limits Results
Averaged Mode DTMA Channel 1 Channel 2 Channel 3 Channel 4
Maximum (ps) 1998.0 ___________ ___________ ___________ ___________
Minimum (ps) 1994.0 ___________ ___________ ___________ ___________
Std. Deviation 0.38 ___________ ___________ ___________ ___________
Non-Averaged Mode DTMA Channel 1 Channel 2 Channel 3 Channel 4
Maximum (ps) 2023.0 ___________ ___________ ___________ ___________
Minimum (ps) 1969.0 ___________ ___________ ___________ ___________
Std. Deviation (ps rms) 5.7 ___________ ___________ ___________ ___________